SpecificationMicrohardness Tester
Test load
A type
mN:
98.07
245.2
490.3
980.7
1,961
2,942
4,903
9,807
gf:1025501002003005001000
Loading mechanismAutomatic load and release method
Load applying speed50
m/sec
Dwell time 5~99sec
Turret mechanismManual
Diamond indenterStandard :Vickers indenter(HV)
Object lens Standard: 2 lenses(x 50 & x10)
Eyepiecex10
Measuring microscope
Type
Electronic
Measuring microscope
Max. measurement length
In case of x 100: 850 m / In case of x 500:
170
m
Measuring microscope
Min. graduation
Digital:0.1
um(In case of x 500)
Measuring microscope
Min. measurement unit
Digital:0.1
m(In case of x 500)
X-Y Stage
Type
Manual
X-Y Stage
Dimensons
100x100mm
X-Y Stage
Max. movement
(X)25x(Y)25mm
X-Y Stage
Min. micro graduation
1/100mm
Precision vise
Max. opening
50mm
Max. height of specimen 95mm
Max. depth of specimen115mm
Availble test parameterHV / HK
Operation panel(sheet sw for type 100 /300 & touch panel sw for type 700)Start/reaset/dwell time/light up-down/turret rotation(for auto-turrettype)
Date display (LED for Type 100/300 & LCD for Type 700)D1/D2/HV-HK/hardness value/loading/OK-NG criteria
Date output & Print outRS232C/centronics(D1/D2/HV-HK/test load/OK-NG)
OK/NG criteriaLimit setting and HI/OK display
Light source 6V/18W
Optical functionAperture diaphragm/Field aperture/Color filter: Replaceable
AccuracyConform to JIS B-7734/B-7725, ASTM E-384, and ISO/DIS 6507-2
Photographic deviceCapable to mount at anytime on the top of the machine (Camera: Option)
Dimensions(W)186x(D)450x(H)504mm
WeightApprox. 40kg(Including standard accessories)
Power supplyJapan: Single phase AC 100V 50/60Hz Overseas: Single phase AC 100~240V 50/60Hz (Should be specified before shipment)